• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS INC. FLEX-NECK®; CATHETER, PERITONEAL, LONG-TERM INDWELLING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS INC. FLEX-NECK®; CATHETER, PERITONEAL, LONG-TERM INDWELLING Back to Search Results
Model Number 00884450330159
Medical Device Problem Codes Fluid/Blood Leak (1250); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that after implanting a peritoneal dialysis catheter in a pediatric patient they experienced issues with leaking.Multiple attempts were made to minimize the leaking; however, they were unsuccessful.Contrast study showed excessive leaking two weeks after implant.The catheter was removed and showed that the catheter had broken into two halves and separated just below the cuff.The patient was treated for exit site infection with oral antibiotics and a new device was implanted.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The complaint was confirmed; however, the root cause could not be determined.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEX-NECK®
Common Device Name
CATHETER, PERITONEAL, LONG-TERM INDWELLING
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS INC.
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
bryson heaton
1600 merit parkway
south jordan, UT 84095
8012084662
MDR Report Key15254803
Report Number1721504-2022-00073
Device Sequence Number6664016
Product Code FJS
UDI-Device Identifier00884450330159
UDI-Public00884450330159
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K970159
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number00884450330159
Device Catalogue NumberCF-5250/C
Device Lot NumberH2203195
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date11/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-