MAUDE Adverse Event Report: STRYKER GMBH LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIIXATION, INTRAMEDULLARY AND ACCESORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIIXATION, INTRAMEDULLARY AND ACCESORIES Back to Search Results
Model Number 1896-5050S
Medical Device Problem Codes Device Difficult to Setup or Prepare (1487); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.The device disposition is unknown.
 
Event or Problem Description
It was reported that there was a failure with the t2 femur screw 5,5mm.The screw didn't rotate in the nail.
 
Event or Problem Description
It was reported that there was a failure with the t2 femur screw 5,5mm.The screw didn't rotate in the nail.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned, and no additional information was available.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If any additional information is provided, the investigation will be reassessed.
 
Additional Manufacturer Narrative
Please note the corrections made to the h6 device code and the manufacturer entity: the reported event was not confirmed, since the device was not returned, and no additional information was available.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If any additional information is provided, the investigation will be reassessed.
 
Event or Problem Description
It was reported that there was a failure with the t2 femur screw 5,5mm.The screw didn't rotate in the nail.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM
Common Device Name
ROD, FIIXATION, INTRAMEDULLARY AND ACCESORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15255177
Report Number0009610622-2022-00381
Device Sequence Number6636026
Product Code HSB
UDI-Device Identifier04546540202512
UDI-Public04546540202512
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K203819
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model Number1896-5050S
Device Catalogue Number18965050S
Device Lot NumberK0F337E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/25/2022
Supplement Date Received by Manufacturer10/13/2022
09/28/2023
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date10/26/2022
10/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/10/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-