ABBOTT VASCULAR TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012274-25
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Above rated burst pressure.The device was discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event or Problem Description
It was reported that the procedure was to treat a mildly tortuous, 50-70% stenosed lesion in the left anterior descending (lad) artery.A 3x25mm trek balloon dilatation catheter (bdc) was advanced to the target lesion for pre-dilatation and inflated one time to 20 atmospheres (atm) when the balloon ruptured.Some of the balloon pieces were snared from the anatomy and other pieces could not be snared and were stented in place in the anatomy.After the procedure the patient seemed fine and was discharged from the hospital that same day.There was no adverse patient sequela.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It was reported that the balloon dilatation catheter (bdc) was overinflated to 20 atmospheres (atms) and the balloon ruptured.It should be noted that the coronary dilatation catheters (cdc), trek rx global, instruction for use (ifu) states: balloon pressure should not exceed the rated burst pressure (rbp).The rbp for the trek rx device is 14 atm; therefore, rbp was exceeded.In this case, it is likely the ifu deviation resulted in the reported balloon rupture.The investigation determined the reported balloon rupture appears to be related to user error; however, the reported unexpected medical intervention, device embedded in tissue or plaque and removal of foreign body appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15255484
Report Number2024168-2022-08921
Device Sequence Number6672246
Product Code LOX
UDI-Device Identifier08717648138454
UDI-Public08717648138454
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model Number1012274-25
Device Catalogue Number1012274-25
Device Lot Number00429G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer08/25/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/29/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexUnknown
Patient Weight112 KG
-
-