• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DIB00
Medical Device Problem Codes Break (1069); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the intraocular lens (iol) cracked when trying to load it.Reportedly, the loading issue occurred during the handling of the product.There was no patient contact with the device.No further information was provided.
 
Additional Manufacturer Narrative
If implanted give date: not applicable as the lens was not implanted.If explanted give date: not applicable as the lens was not explanted.Initial reporter email address: unknown/not provided.The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, it was not available.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Additional Manufacturer Narrative
Section a2 age/date of birth: date of birth is (b)(6) 1952 and age is 70.Section a3 gender: female.Section d9: device available for evaluation? yes.Section d9: date returned to manufacturer: sep 6, 2022.Section h3: evaluated by manufacturer: yes device evaluation: visual inspection under magnification of the complaint handpiece revealed that it was received with the plunger rod advanced into the cartridge.The cartridge tip could be observed to be cracked and the cartridge neck could be observed to be damaged.The handpiece was disassembled and the assembly was inspected, no assembly issues were identified.The complaint lens was removed from the tape and cleaned.Revealing that the edge of the lens was chipped.A similar chip could also be identified on the haptic.A product quality deficiency could not be determined.The complaint issue " dc-lens damaged" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Conclusion: as a result of the investigation there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15255732
Report Number3012236936-2022-02064
Device Sequence Number6720801
Product Code HQL
UDI-Device Identifier05050474731684
UDI-Public(01)05050474731684(17)240423
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2024
Device Model NumberDIB00
Device Catalogue NumberDIB00U0165
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer09/25/2022
Initial Report FDA Received Date08/18/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-