• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER DENTAL IMP, TSV, 3.7, 10,MTX,MC,MG,HA; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER DENTAL IMP, TSV, 3.7, 10,MTX,MC,MG,HA; DENTAL IMPLANT Back to Search Results
Model Number TSVMH10
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant was removed due to becoming fractured at the site.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).Patient weight not provided.Event date not provided.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).An implant was returned for investigation.Visual inspection of the as returned product identified worn marking due to usage, bone and a fracture at the collar.Also, an associated product hc343 has not been returned.However, this item is listed as associated item only and did not cause or contribute to the event.No further investigation will be conducted.No pre-existing conditions were noted on the per.The reported device was location is 18 and was used for approximately 6 years.Pictures or x-ray images were not provided.Review of appropriate documentation: instructions for use for tapered screw-vent® and trabecular metal¿ implants ifu 4869 rev 9 ¿ 10/19 information identified: 'contraindications' 'warnings' 'changes in performance' 'adverse effects' dhr review: dhr review was completed for the subject lot number (63802111).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (63802111) for similar event and no other complaint was identified.Post market trend review: nov post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur and the reported event was confirmed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP, TSV, 3.7, 10,MTX,MC,MG,HA
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMMER DENTAL
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15257536
Report Number0002023141-2022-02111
Device Sequence Number14905710
Product Code DZE
UDI-Device Identifier00889024344099
UDI-Public(01)00889024344099(10)63802111(17)221031
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K133339
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/31/2022
Device Model NumberTSVMH10
Device Catalogue NumberTSVMH10
Device Lot Number63802111
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer01/17/2023
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date01/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexFemale
-
-