• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM PREC NEURO MATCH HEAD; DRILLS, BURRS, TREPHINES & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP 3.0MM PREC NEURO MATCH HEAD; DRILLS, BURRS, TREPHINES & ACCESSORIES Back to Search Results
Model Number 5820107530
Medical Device Problem Codes Fitting Problem (2183); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
 
Event or Problem Description
It was reported that prior to a procedure, the bur would not fit the attachment, it fell out.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
 
Additional Manufacturer Narrative
Corrected data: b5: it was initially reported that the bur fell out of the attachment, upon evaluation of the returned device, the notch at the mount end was incomplete and could not fit into the attachment.This is not a reportable event.H6: the quality investigation is complete.
 
Event or Problem Description
It was reported that prior to a procedure, the bur would not fit the attachment, it fell out.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.Update: the returned device had an incomplete notch and could not fit into the attachment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM PREC NEURO MATCH HEAD
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES
Manufacturer (Section D)
STRYKER INSTRUMENTS-A DIVISION OF STRYKER CORP
1941 stryker way
portage MI 49002
Manufacturer (Section G)
STRYKER INSTRUMENTS-IRELAND
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
Manufacturer Contact
una barry
instruments division
carrigtwohill bus. & tech park
carrigtwohill NA
EI   NA
214532900
MDR Report Key15257764
Report Number3015967359-2022-01655
Device Sequence Number15125345
Product Code HBE
UDI-Device Identifier07613154028177
UDI-Public07613154028177
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K143320
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5820107530
Device Catalogue Number5820107530
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-