| Model Number |
5820107530 |
| Medical Device Problem Codes |
Fitting Problem (2183); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow up report will be filed once the quality investigation is complete.Awaiting device return.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a procedure, the bur would not fit the attachment, it fell out.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b5: it was initially reported that the bur fell out of the attachment, upon evaluation of the returned device, the notch at the mount end was incomplete and could not fit into the attachment.This is not a reportable event.H6: the quality investigation is complete.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to a procedure, the bur would not fit the attachment, it fell out.It was also reported that this event did not occur during a surgical procedure and there was no delay or adverse consequences as a result of this event.Update: the returned device had an incomplete notch and could not fit into the attachment.
|
| |
|