MAUDE Adverse Event Report: SYNTHES GMBH DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number 05.001.083
Medical Device Problem Codes Material Rupture (1546); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.Udi: (b)(4).
 
Event or Problem Description
It was reported from germany that during service and evaluation, it was determined that the air hose device hose ruptured, had an air leak and failed pre-test for external pressure test.It was noted in the service order that the device pipe separates from the nozzle.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key15257798
Report Number8030965-2022-05866
Device Sequence Number5787518
Product Code HBB
UDI-Device Identifier10886982129048
UDI-Public10886982129048
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number05.001.083
Device Catalogue Number05.001.083
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Initial Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-