| Model Number |
9734685 |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: 9733763, version#: 2.2.7.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used intraoperatively during a functional endoscopic sinus surgery (fess).It was reported that the site registered multiple times and couldn't get an accurate registration.The use of medtronic navigation was aborted.There was a surgical delay of less than one hour.There was no impact on the patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: a software analysis was initiated.However, it was found that there was insufficient information to determine the relationship of the software to the reported issue.The logs captured one session on the date of the issue and recorded multiple registration attempts by the site.Archives consists of 5 ct exams.Ct-5 and ct-3 with 195 slices and 244 slices were the non-axial scans loaded with a block 3d model (skin model is achieved by changing the threshold), ct-3 with 181 slices loaded a clear 3d model but none of the exams had the registration data to check the trace pattern.H6: b01, c19 and d15 apply to the software analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|