Model Number AMSCO 400 STEAM STERILIZER
Medical Device Problem Codes Mechanical Jam (2983); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user facility reported via medwatch #(b)(4) that the door locking mechanism on their amsco 400 sterilizer would not lock or unlock causing loads to be stuck inside or cause loads to not be run.No procedure delays reported.
 
Additional Manufacturer Narrative
A steris service technician inspected the sterilizer and found the hinged door plate required replacement.Investigation of the reported event is currently in process.A follow-up report will be submitted when additional information becomes available.
 
Additional Manufacturer Narrative
The steris service technician replaced the hinge door plate, tested the sterilizer, and found the unit to be operational.The sterilizer was returned to service and no additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMSCO 400 STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15257844
Report Number3005899764-2022-00037
Device Sequence Number12198781
Product Code FLE
UDI-Device Identifier00724995210137
UDI-Public00724995210137
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAMSCO 400 STEAM STERILIZER
Device Catalogue NumberSR2221211011
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-