• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 52MM; PROSTHESIS, HIP, SEMI-CONSTR, METAL/POLY, CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 20 DEG XLPE ACET LNR 36MM X 52MM; PROSTHESIS, HIP, SEMI-CONSTR, METAL/POLY, CEMENT Back to Search Results
Model Number 71335752
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal reference (b)(4).
 
Event or Problem Description
It was reported that after a thr performed on (b)(6) 2022, the patient had an infection.A revision surgery was addressed to explant all the implants (oxonium fem hd 12/14 32 mm +4, r3 20 deg xlpe acet lnr 36 mm x 52 mm, r3 0 hole ha ctd acet shell 52 mm, ref interfit thrd hole cover and polarstem stem lat.Ti/ha 1 non-cem) and insert cement spacers.Further information is unknown at the moment.
 
Additional Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the infection.It is noted the infection was addressed via revision surgery to explant all the implants and insert cement spacers.The patient impact beyond the reported revision cannot be determined with the information provided.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that infection, both early, post-operative superficial and early, post-operative deep wound infection and late periprosthetic infection has been identified in potential complications associated with total hip arthroplasty surgery, primary or revision section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to loss of sterility during procedure, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 20 DEG XLPE ACET LNR 36MM X 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTR, METAL/POLY, CEMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15257969
Report Number1020279-2022-03758
Device Sequence Number6671994
Product Code JDI
UDI-Device Identifier03596010598332
UDI-Public03596010598332
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K113848
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71335752
Device Catalogue Number71335752
Device Lot Number21FM22940
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2022
Supplement Date Received by Manufacturer09/28/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-