EXACTECH, INC. TRULIANT TIB IMP PS INSERT SZ 2.5 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
|
Back to Search Results |
|
| Model Number |
TRULIANT TIB IMP PS INSERT SZ 2.5 11MM |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Unstable (1667)
|
| Health Effect - Clinical Code |
Joint Laxity (4526)
|
| Date of Event |
07/18/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 69 yo female patient, who was initially implanted on the right knee on (b)(6) 2021, underwent a poly swap revision surgery approximately 1 year 1 month post the initial implant procedure due to instability.They increased poly thickness from a 11mm ps insert to a 15mm psc insert and were revised to exactech devices.Patient was last known to be in stable condition following the event.Due to hospital policies, no products are available for return.No additional information is available.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 02-020-11-0325 - truliant ps cem fem ps cem right sz 2.5, 6792590; 02-022-45-2525 - truliant tib fit tray cem sz 2.5f / 2.5t, 6531018; 200-02-35 - three peg patella 35mm, 6470897; 201-78-15 - holding pin mini sharp point 4 pk, 6790703; 521-78-23 - threaded pin size 2.3 collared 2pk, s265617; 521-78-31 - threaded pin size 2.6 collarless 2pk, s110679.No device returning due to hospital policies.Recall: z-2112 thru 2133-2021.
|
| |
|
Additional Manufacturer Narrative
|
|
H7: z-0023-2022.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, d9, h6: medical device problem code, component code, type of investigation, investigation findings, and investigation conclusions.Based on review of all available information, there is no evidence to suggest that the reported event is related to any design or manufacturing issues.The cause of the subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.These devices are used for treatment not diagnosis.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
|
| |
|