| Model Number |
10491449 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reported a discrepant low thb (total hemoglobin) result generated on the rp500 compared to the thb result of a different sample on a non-siemens instrument.There is no report of injury due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway which includes a review of the instrument log files the customer provided.Customer is operational.The cause of this event is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Siemens has completed the investigation.The definitive root-cause for the low thb could not be determined.Based on the investigation this rp500 system appeared stable and functioning as expected.Data did not suggest anything unusual about the co-ox behavior of this patient or that hemodialysis impacted the thb result.The reported thb results appear valid.For an accurate measurement of thb, thorough mixing and multi-directional rotating of the red blood cells immediately before analysis is required.This is particularly important for blood samples of lower hematocrit (red blood cells) as the cells more quickly settle out, falling out of homogeneity.Pre-analytical sample condition(s) is most likely a primary contributor to the observed discordant thb values.Cause of this event is unknown.
|
| |
|