| Model Number |
21-2863-24 |
| Medical Device Problem Codes |
Failure to Advance (2524); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that when priming the product during the pre-use check, he felt an unusual resistance through it and was unable to advance the syringe.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: h6: health impact and evaluation codes: updated.H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found no damage can be observed on the base and needle.Functional testing found occlusion was observed between the tube and needle base.The complaint is confirmed.The root cause of the reported issue was found to be manufacturing - excess solvent.Awareness notification was made to production personnel in order to explain the importance to adherence or following in the procedure ".Gripper huber needle luer bonding assembly.Section 5.0 solvent bond tube to luer by quality engineer on november 04, 2022.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Search Alerts/Recalls
|