Model Number 21-2863-24
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product sample was received and is awaiting evaluation and investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that when priming the product during the pre-use check, he felt an unusual resistance through it and was unable to advance the syringe.No patient injury was reported.
 
Additional Manufacturer Narrative
Other text: h6: health impact and evaluation codes: updated.H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing was performed.Visual inspection found no damage can be observed on the base and needle.Functional testing found occlusion was observed between the tube and needle base.The complaint is confirmed.The root cause of the reported issue was found to be manufacturing - excess solvent.Awareness notification was made to production personnel in order to explain the importance to adherence or following in the procedure ".Gripper huber needle luer bonding assembly.Section 5.0 solvent bond tube to luer by quality engineer on november 04, 2022.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC GRIPPER PLUS NEEDLE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
7-1-1 akasaka,
minneapolis, MN 55442
MDR Report Key15258676
Report Number3012307300-2022-15657
Device Sequence Number17579045
Product Code FPA
UDI-Device Identifier10610586025679
UDI-Public10610586025679
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number21-2863-24
Device Catalogue Number21-2863-24
Device Lot Number4232500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2022
Supplement Date Received by Manufacturer12/22/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-