• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLINISYS INC SUNQUEST LABORATORY - BLOOD BANK AND BLOOD DONOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLINISYS INC SUNQUEST LABORATORY - BLOOD BANK AND BLOOD DONOR Back to Search Results
Model Number 10.0 AND ABOVE
Medical Device Problem Code Loss of Data (2903)
Health Effect - Clinical Code Unspecified Blood or Lymphatic problem (4434)
Date of Event 08/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
39 sites have been notified by a client notification and a correction will be available in blood bank versions 10.0.1, 10.1.1 and 11.0.1.There are 0 sites that have already been corrected.No instances of patient harm have been reported as a result of this issue.The root cause of this problem is coding logic.When definitions exceed 497 entries the routine handling the maintenance does not look in the correct database location (^tpo(dv,"pkt")) for the definitions.The routine is expecting a variable that no longer contains the information and it processes the definitions as if they all have been removed.
 
Event or Problem Description
When new gui blood type compatibilities maintenance definitions are saved to the database, if the number of overall compatible and suitable definitions exceeds 497 lines of data, all definitions are removed.If maintenance is removed, and problem code definitions were defined, patient/unit blood type compatibilities based on problem code definitions will not be generated.This could result in a patient receiving incompatible blood.Maintenance for patient/unit blood type compatibility (bma 10,2) are not affected by this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUNQUEST LABORATORY - BLOOD BANK AND BLOOD DONOR
Common Device Name
BLOOD BANK
Manufacturer (Section D)
CLINISYS INC
3300 e. sunrise dr.
tucson AZ 85718
Manufacturer (Section G)
CLINISYS INC
3300 e. sunrise dr.
tucson AZ 85718
Manufacturer Contact
chris harmston
3300 e. sunrise dr.
tucson, AZ 85718
7206603377
MDR Report Key15258905
Report Number2029302-2022-00001
Device Sequence Number12215132
Product Code MMH
UDI-Device IdentifierB236BB10000
UDI-Public+B236BB10000/$$710000063
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
BK160005
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Modification/Adjustment
Type of Report Initial
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number10.0 AND ABOVE
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/22/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-