BOSTON SCIENTIFIC CORPORATION REZUM; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number G2200
Medical Device Problem Code Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that following sedation of the patient, this rezum generator displayed a red light and message to call support/replace the unit.An unspecified error message was observed and restarting the generator was attempted but did not resolve the issue.As a result, the procedure was canceled to be rescheduled on another date.No patient complications were reported.
 
Event or Problem Description
It was reported that following sedation of the patient, this rezum generator displayed a red light and message to call support/replace the unit.An unspecified error message was observed and restarting the generator was attempted but did not resolve the issue.As a result, the procedure was canceled to be rescheduled on another date.No patient complications were reported.
 
Additional Manufacturer Narrative
Upon receipt, the rezum generator was thoroughly inspected and analyzed.Courtesy message 210 (temperature reading is above 590 degrees c) was confirmed to be present and reproducible.The internal device cable was replaced, and the issue resolved.The generator additionally received a plastic enclosure upgrade.The generator passed all functional and electrical safety testing.Based on the results of this investigation, the most probable cause of the event was determined to be the result of the observed component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REZUM
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15259663
Report Number2124215-2022-31497
Device Sequence Number6688072
Product Code GEI
UDI-Device Identifier08714729986850
UDI-Public08714729986850
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K180237
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG2200
Device Catalogue NumberG2200
Device Lot Number0000003271
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer02/28/2023
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date03/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2017
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-