Patient reported, "experiencing pain".Healthcare professional reported, left side rupture.And textured to smooth exchange, due to patient concern.Device has been explanted and replaced.
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: pain, concern with textured product.Device evaluation: analysis of the returned device identified, underweight, broken shell and missing shell.A visual and microscopic analysis was performed which identified, sharp broken on anterior assessed, as a surgical impact opening, for the stress mark in the shell, crease fold and wear abrasion.A dimension measurement in the shell was performed which identify, the thickness within specification.Based on the device analysis, the final assessment is: a sharp broken on anterior assessed, as a surgical impact opening.Missing shell assessed, as inconclusive.
|