The reported event could be confirmed, since the device was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the received screw was found to be broken from the mid-shaft region.A few threads were found to be deformed around the nail contact region, indicating towards high compressive load.Presence of lines of rest and a smooth topography indicates the fracture to be in a fatigue manner due to high cyclic loading.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.A clinical opinion regarding the event was sought from an independent medical professional based on the available patient information and the given x-ray.Question regarding breakage was asked to which he responded: "there are patient related factors (the fracture pattern with a subtrochanteric fracture just below the lesser trochanter) -other information is difficult to assess from only one plane (there is the axial plane missing and pre-or at least directly postoperative x-ray would be helpful, here!).Surgical factors also could have influenced the failure (miss-drilling can reduce the resistance to load significantly).We see a hypertrophic non-union (this may indicate too much micromotion at the fracture site).Therefore the scenario would be: misdrilling led too reduced resistance to load, micromotion led to hypertrophic non-union, finally the fracture did not consolidate and failure with breakage occurred." based on the above investigation and available information, the root cause of the failure is deemed to be both user and patient related.Appearance of mis-drilling marks indicates that the nail was damaged intra-operatively which may have led to weakening of the nail initially.Further to this, non-union and the simultaneous overloading on the whole construct ultimately led to the breakage of the nail and subsequently a distal locking screw as well, in a fatigue manner.The instruction for use clearly instructs the user that: ¿these implants are neither intended to carry the full load of the patient acutely, nor intended to carry a significant portion of the load for extended periods of time.For this reason post-operative instructions and warnings to patients are extremely important.External immobilization (e.G.Bracing or casting) may be employed until x-rays or other procedures confirm adequate bone consolidation.[¿] these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in a position to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device.¿ [original statement(s)] if any additional information is received, the record will be updated.
|