• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1896-5050S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Nonunion/Delayed-union Bone Fracture (2369)
Date of Event 05/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
The customer reported a broken gamma3 nail (3525-0360s)."the patient was treated externally for a per/subtrochanteric left femur fracture on (b)(6) 2021 with a with a gamma nail.During the consultation (b)(6) 2022, the implant fracture was diagnosed and the patient was admitted to hip clinic at balgrist university hospital for further treatment.Here the indication for revision was given in case of symptomatic pseudarthrosis.The indication for revision was given and the patient had to undergo surgery on (b)(6) 2022." 7/22/2022 - additional information: upon product return, one locking screw was found broken.
 
Event or Problem Description
The customer reported a broken gamma3 nail (3525-0360s)."the patient was treated externally for a per/subtrochanteric left femur fracture on (b)(6) 2021 with a with a gamma nail.During the consultation (b)(6) 2022, the implant fracture was diagnosed and the patient was admitted to hip clinic at (b)(6) hospital for further treatment.Here the indication for revision was given in case of symptomatic pseudarthrosis.The indication for revision was given and the patient had to undergo surgery on (b)(6) 2022." 7/22/2022 - additional information: upon product return, one locking screw was found broken.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the received locking screw is broken apart as complained.The device inspection revealed the following: the visual inspection has shown that the locking screw is broken apart twice, the front fragment of the screw was not returned for evaluation.The thread flanks of the screw were partially flattened in the area where the screw was in contact with the nail, which indicates that the screw was exposed to high loads over a longer period of time in this area.Under the microscope the typical characteristics of a fatigue fracture can be identified at the fracture face.The origin of the crack is visible, there is a fatigue zone with a smooth surface and progression lines over almost the whole surface of the fracture and then we have a shear lip, where the final fracture occurred.The shiny surface on one side indicates that the fragments rubbed against each other after the breakage.The relevant dimensions were verified during the investigation and no deviation from the specification could be detected.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a patient related issue.The reported pseudarthrosis did finally lead to a fatigue failure of the implants.In this relation following statement from the instructions for use can be mentioned: ¿these devices can break when subjected to the increased loading associated with delayed unions and/or non-unions.Internal fixation devices are load sharing devices which are intended to hold fractured bone surfaces in apposition to facilitate healing.If healing is delayed or does not occur, the appliance may eventually break due to metal fatigue.Loads on the device produced by load bearing and the patient¿s activity level will dictate the longevity of the device." if updated information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW, FULLY THREADED T2 TIBIA Ø5X50 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15260489
Report Number0009610622-2022-00385
Device Sequence Number14899841
Product Code HSB
UDI-Device Identifier04546540202512
UDI-Public04546540202512
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K032244
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model Number1896-5050S
Device Catalogue Number18965050S
Device Lot NumberK0542A0
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/20/2022
Supplement Date Received by Manufacturer09/20/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date10/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-