The initial reporter stated they received questionable values for two samples from the same patient tested with elecsys anti-hcv ii.The first complained sample (sample 3) was tested on two cobas 8000 e 801 module analyzers.The second complained sample (sample 4) was tested on an unknown roche analyzer.The sample results were reported outside of the laboratory.On (b)(6) 2022, a first sample collected from the patient was tested using the abbott alinity hepatitis c method, resulting in a value of 3.83 s/co (reactive).On (b)(6) 2022, a second sample collected from the patient was tested using the abbott real time hcv viral load quantification method, resulting in a value of 5311136 iu/ml (6.72 log).On (b)(6) 2022, a third sample collected from the patient was tested using the elecsys anti-hcv on the first e 801 analyzer, resulting in a value of 0.0439 coi (non reactive).This sample was repeated on a second e 801 analyzer on (b)(6) 2022, resulting in a non-reactive value.On (b)(6) 2022, a fourth sample collected from the patient on (b)(6) 2022 was tested on an unknown roche analyzer on (b)(6) 2022, resulting in a value of 0.045 coi (non reactive).On (b)(6) 2022, a fifth sample collected from the patient was tested with the cobas hcv pcr viral load/hcv pc quantitative kit, resulting in a value of 1400000 iu/ml (7.15 log).On (b)(6) 2022, a sixth sample collected from the patient was tested using the riba anti hcv- immunoblot 3rd generation method, resulting in the following: conclusion: indeterminate core: negative ns3-1: 3+ ns3-2: negative ns4: negative ns5: negative.On (b)(6) 2022, a seventh sample collected from the patient was tested twice on the second e 801 analyzer, resulting in values of 1.27 coi (reactive) and 1.22 coi (reactive).A serial number of the first e 801 analyzer is (b)(4).It was asked, but it is not known what the serial numbers were for the second e 801 analyzer or unknown roche system.Sample 3 was tested using elecsys anti-hcv reagent lot 58750101, with an expiration date of 30-nov-2022.Sample 4 was tested using elecsys anti-hcv reagent lot 587906.The expiration date for this lot was requested, but not provided.
|