Model Number TB-0520IC
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that during a therapeutic laparotomy procedure, the tissue pad was peeling off.The procedure was completed by using another device of the same model.There was no health hazard to the patient.Additional details have been requested regarding the reported event.At this time, no additional information has been provided.
 
Additional Manufacturer Narrative
The subject device is expected to be returned to olympus for evaluation, but has not yet been received.The investigation is ongoing; therefore, a root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
Correction to g2, health professional was inadvertently selected on the initial report.This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated where the reported event was not confirmed.No peeling of the tissue pad was confirmed however, the tissue pad was noted to be worn.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the tissue pad was worn because the grip was closed and ultrasonic waves were output while nothing was gripped between the grip and the tip of the probe (including after the tissue was cut).The following is included in the instructions for use (ifu): "do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating." "when cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation" olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 20 CM, INLINE GRIP
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15262247
Report Number9614641-2022-00168
Device Sequence Number6688160
Product Code GEI
UDI-Device Identifier04953170308710
UDI-Public04953170308710
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132703
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTB-0520IC
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date08/19/2022
Supplement Report FDA Received Date10/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-