• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE SPINE, INC. PROLIFT; INTERVERTEBRAL BODY FUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE SPINE, INC. PROLIFT; INTERVERTEBRAL BODY FUSION DEVICE Back to Search Results
Model Number L151-596
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Installer was not returned, based on the potential failure per given description of the issue, it is hypothesized that during excessive force was applied to the installe at an angle which reulted in a torsional or off-axis asymmetric (rocking or toggling) load being applied to the distal tip of the installer.
 
Event or Problem Description
Installer broke off during surgery, a piece was left in the patient but surgeon was able to retrieve the broken portion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROLIFT
Common Device Name
INTERVERTEBRAL BODY FUSION DEVICE
Manufacturer (Section D)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer (Section G)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer Contact
angela batker
13951 south quality drive
huntley, IL 60142
8478846117
MDR Report Key15262619
Report Number3004499989-2022-00009
Device Sequence Number17551267
Product Code MAX
UDI-Device Identifier00190837117610
UDI-Public00190837117610
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K190488
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberL151-596
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-