• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION RENEGADE STC 18; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION RENEGADE STC 18; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 80617
Medical Device Problem Codes Detachment of Device or Device Component (2907); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device tip was broken.The target lesion was located in the moderately tortuous internal iliac artery.A 130/20 renegade stc-18 catheter infusion was selected for use.During the procedure, the device was hydrated using intermittent flushing.Upon advancing, the coil cannot be inserted, and it was noted that the tip was broken and crushed.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.The hub and shaft were visually inspected when returned.The renegade device showed multiple bends and kinks on the distal end of the device.Per the customers alleged complaint that the coil would not go through the device, the device was x-rayed to see if the coil was stuck in the lumen.No coil was noticed.A.014 guidewire was inserted into the device and the wire transcended through the stc- 18 with no hesitations or restrictions.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.The complaint was not confirmed for coils in the catheter; however, shaft damage was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENEGADE STC 18
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15262739
Report Number2124215-2022-28719
Device Sequence Number14928055
Product Code DQO
UDI-Device Identifier08714729444442
UDI-Public08714729444442
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K023681
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/15/2024
Device Model Number80617
Device Catalogue Number80617
Device Lot Number0028023873
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Initial Report FDA Received Date08/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-