• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION R2P¿ DESTINATION SLENDER¿ GUIDING SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION R2P¿ DESTINATION SLENDER¿ GUIDING SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual device was not returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot number combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that the destination slender guiding sheath sprayed contrast on to physician and technologist upon hand injecting contrast through the sheath.The sheath valve sprays contrast and could also blood can spray all over.The patient was stable, and the procedure was successful.Additional information was received on (b)(6) 2022: the type of procedure being performed was a radial to peripheral angioplasty with stent placement.The procedure was completed using the device.Sprays from the valve upon hand injecting contrast may have come out from leaflets that support the wire in center of sheath.There was no noticeable damage to the device.There was no blood loss.
 
Additional Manufacturer Narrative
This report is being submitted as follow-up no.1 to provide the completed investigation results.The actual device was not returned; therefore, an evaluation of the actual device was unable to be conducted.The complaint cannot be confirmed since the device was not available for evaluation.The exact root cause cannot be determined.The likely root cause is off-center insertion of the dilator through the sheath valve.The device history record (dhr) review determined that the device was in a conforming state when released from terumo control.There is no indication that any manufacturing issues may have led to this event.Currently no action is recommended since this risk evaluation is within the predetermined limits in the failure mode & effects analysis (fmea).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R2P¿ DESTINATION SLENDER¿ GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer (Section G)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
Manufacturer Contact
gina digioia
950 elkton blvd.
elkton, MD 21921
6402040886
MDR Report Key15263499
Report Number1118880-2022-00062
Device Sequence Number20336782
Product Code DYB
UDI-Device Identifier00389701012353
UDI-Public00389701012353
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
K193125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberGS-R6ST1C10W
Device Lot Number0000106156
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date08/21/2022
Supplement Report FDA Received Date09/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-