• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO MALE CATHETER WITH SUREGRIP; INTERMITTENT CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3 GO MALE CATHETER WITH SUREGRIP; INTERMITTENT CATHETER Back to Search Results
Model Number 53812G
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Device Handling Problem (3265)
Health Effect - Clinical Code Urinary Tract Infection (2120)
Date of Event 08/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event or Problem Description
It was reported that the patient got a urinary tract infection after using male magic 3 go catheter.Now patient was doing a return for female magic3 go intermittent catheter (51812 catheters).It was unknown what medical intervention was provided for urinary tract infection.
 
Event or Problem Description
It was reported that the patient got a urinary tract infection after using male magic 3 go catheter.Now patient was doing a return for female magic3 go intermittent catheter (51812 catheters).It was unknown what medical intervention was provided for urinary tract infection.
 
Additional Manufacturer Narrative
The reported event was confirmed as use related as the female patient used the male catheter.A potential root cause for this failure could be "lack of existing patient/caregiver/physician education/physician screening, user error".It was unknown whether the device had met specifications.The product was used for treatment purposes.The failure was caused by the misuse of the product.A dhr review was not required because the investigation was confirmed as use related.The instructions for use were found adequate and state the following: "inspect the catheter before use.Do not use the product if the device or packaging is damaged.Self-catheterization should follow the plan of care and advice given by your healthcare practitioner and be carried out only in accordance with the instructions for use provided.Please consult your healthcare practitioner if any conditions occur which create concern and/or difficulty during catheterization." corrections: f,h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3 GO MALE CATHETER WITH SUREGRIP
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15263983
Report Number1018233-2022-06585
Device Sequence Number6599401
Product Code EZD
UDI-Device Identifier00801741136955
UDI-Public(01)00801741136955
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K172247
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number53812G
Device Catalogue Number53812G
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer09/27/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date09/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexFemale
-
-