ABBOTT DIABETES CARE INC XTRA METER 3 BUTTON PRECISION; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 98814-65
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
To date, product has not been returned for further investigation.The useful life of precision xceed meter is 5 years.As the manufacturing date of this product is 2016, it has been in distribution beyond its useful life at the time of the complaint.Since the product exceeded its useful life, it is determined to have met specification when the product was released and through its lifespan.No further investigation activities are being performed at this time.All pertinent information available to abbott diabetes care has been submitted.
 
Additional Manufacturer Narrative
The returned meter was investigated with retained test strips.Visual inspection was performed on returned meter.No issues observed.Meter powered on with button depression and strip insertion.Blank screen was not observed.The complaint was not confirmed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTRA METER 3 BUTTON PRECISION
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key15264409
Report Number2954323-2022-30223
Device Sequence Number5871281
Product Code NBW
UDI-Device Identifier00093815998372
UDI-Public00093815998372
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K040814
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number98814-65
Device Catalogue Number99837
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/15/2022
Supplement Date Received by Manufacturer10/04/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
Patient Weight122 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-