ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO MED/LG APPLIER Back to Search Results
Catalog Number EL314
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an unknown procedure the clip did not close securely and in line.
 
Additional Manufacturer Narrative
(b)(4).Batch # unk.Date of event: only event year known: 2022.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Photo analysis: this is an analysis of two photos submitted to ethicon for review.During the visual analysis the following was observed: the photographs show four malformed clips and the tip of a ligaclip*endo med/lg applier device, the picture is not clear to determine the failure mode.Based on the photo review, the event describe is confirmed, however no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the instrument was not returned our evaluation is limited.We value the opportunity to fully analyze the instrument upon its return.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.The manufacturing records could not be reviewed as the batch number is unknown.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: please provide more details: was there any bleeding? if yes, how was the bleeding controlled? what amount of blood loss (mls) occurred? was a transfusion required? was there any change to the procedure as a result of the event? were there any patient consequences? if yes, please describe.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
LIGACLIP*ENDO MED/LG APPLIER
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
329348013
MDR Report Key15264793
Report Number3005075853-2022-05470
Device Sequence Number17792099
Product Code HBT
UDI-Device Identifier10705036003854
UDI-Public10705036003854
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K830503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEL314
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-