| Catalog Number |
8065000128 |
| Medical Device Problem Codes |
Complete Blockage (1094); Noise, Audible (3273); Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported tubing clogging and system occlusion tone was heard during irrigation/aspiration phase of a cataract surgery.The product was replaced and the surgery was completed.There was no patient harm.
|
| |
|
Event or Problem Description
|
|
Upon further review of the follow up information there was loss of suction and it would not remove ophthalmic viscosurgical device (ovd) during irrigation/aspiration phase of a cataract surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.No sample has been returned for evaluation; therefore, the condition of the product could not be verified.Visual inspection or functional testing could not be conducted in order to ascertain the failure mode for the consumable device.The root cause of the customer's complaint could not be established as a sample has not been received and the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known, therefore, specific action with regards to this complaint cannot be taken.Based on our current tracking, there are no adverse trends for this reported complaint.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|