MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Failure to Deliver Energy (1211); Migration or Expulsion of Device (1395); Energy Output Problem (1431); Application Program Problem (2880); Battery Problem (2885); Data Problem (3196)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/14/2022 |
Event Type
malfunction
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for non-malignant pain.The reason for call was caller stated that a couple of days ago they started getting a message that said "settings not available." caller stated they aren't able to increase their desired settings and usually the intensity is set in the 4s but now it's down to 2.3.Caller stated that last night they were making their way home and the implant briefly lost power and ever since recharging it, they can't get it to turn back on and it doesn't seem like the stimulation is on.Caller added that they are able to adjust settings within groups b and c, but those groups are not as effective as group a where they're seeing the message.The patient was redirected to their healthcare provider/field rep to further address the issue.Caller added that they tried calling their rep but they were on vacation but left the number for another rep who caller was waiting to hear back from.Agent reviewed rep request process if they don't hear back.Redirected caller: patient's hcp.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the patient on 2022-september-09.Pt reported that wire movement led to the issue.A rep changed their settings and the issue was resolved.
|
|
Manufacturer Narrative
|
Continuation of d10: product id 977c265 lot# serial# (b)(6) product type lead medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|