• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TORNIER S.A.S. TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 5C 137.5DEG 82MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 5C 137.5DEG 82MM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Model Number DWF605C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Loss of Range of Motion (2032); Implant Pain (4561)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient sustained work related shoulder injury.The patient underwent surgery to repair the rotator cuff tendon tear.There were ongoing complaints of pain and poor function which led to the patient undergoing a shoulder hemi-arthroplasty.The pain and poor range of movement continued.The patient went to another surgeon for further rotator cuff pathology diagnosis.The surgeon that the patient requires conversion of hemi-arthroplasty to reverse shoulder arthroplasty (rsa).The stem and head implants were removed without difficulty.The stem was fixed.A competitor rsa prosthesis implanted.
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Kept by surgeon.
 
Event or Problem Description
It was reported that a patient sustained work related shoulder injury.The patient underwent surgery to repair the rotator cuff tendon tear.There were ongoing complaints of pain and poor function which led to the patient undergoing a shoulder hemi-arthroplasty.The pain and poor range of movement continued.The patient went to another surgeon for further rotator cuff pathology diagnosis.The surgeon that the patient requires conversion of hemi-arthroplasty to reverse shoulder arthroplasty (rsa).The stem and head implants were removed without difficulty.The stem was fixed.A competitor rsa prosthesis implanted.
 
Additional Manufacturer Narrative
Please note corrections in: b2.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event must be available in order to determine the root cause of the complaint event.But based on the event description which mentions that " further rotator cuff pathology diagnosed" and based on the ifu which mentions that "the population targeted with the anatomic configuration is restricted to patients with an intact or reconstructable rotator cuff, where it is intended to provide increased mobility and stability and to relieve pain.", the probable root cause of this event is probably due to the patient condition.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORNIER FLEX SHOULDER SYS STD HUM PTC STEM 5C 137.5DEG 82MM
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin 38330
FR  38330
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15265158
Report Number3000931034-2022-00273
Device Sequence Number5881104
Product Code KWS
UDI-Device Identifier03700434021479
UDI-Public03700434021479
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K190521
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date07/31/2025
Device Model NumberDWF605C
Device Catalogue NumberDWF605C
Device Lot Number7561AV
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age42 YR
Patient SexMale
-
-