• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #6 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #6 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5510-F-602
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.It was reported the patient is not experiencing any adverse consequences and the surgeon doesn¿t currently plan to revise the patient.However, due to the user¿s incorrect selection this could lead to patient harm, therefore, a serious injury mdr will be filed.
 
Event or Problem Description
Primary procedure, left knee.It was reported that a right-sided femoral component was implanted in the patient's left knee.The implant was shown to o.R.Staff by the rep, who then gave the implant to a nurse, who gave it to the surgical tech, who gave it to the surgeon.Discovery was made after the patient was closed and the surgeon was notified.Surgeon notified the patient's family, who declined a revision of the knee.Surgeon's plan is to observe and assess.Rep provided the usage sheet and confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
An event regarding incorrect selection involving a triathlon femoral component was reported.The event was confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that a right side femoral component was implanted in the patient's left knee instead of a left side femoral component.No patient harm or treatment plan was reported.The reported event was confirmed through review of the provided implant sheet and is considered to be the result of user error as no manufacturing-related product problem was found given the information provided.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Primary procedure, left knee.It was reported that a right-sided femoral component was implanted in the patient's left knee.The implant was shown to o.R.Staff by the rep, who then gave the implant to a nurse, who gave it to the surgical tech, who gave it to the surgeon.Discovery was made after the patient was closed and the surgeon was notified.Surgeon notified the patient's family, who declined a revision of the knee.Surgeon's plan is to observe and assess.Rep provided the usage sheet and confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #6 R-CEM
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
brad curtis
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15265418
Report Number0002249697-2022-01202
Device Sequence Number5881112
Product Code MBH
UDI-Device Identifier07613327040326
UDI-Public07613327040326
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number5510-F-602
Device Catalogue Number5510F602
Device Lot NumberPH73L
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer10/21/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age78 YR
Patient SexMale
-
-