W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Model Number |
TGT3415 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm expansion (4597)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).It should be noted the gore® tag® thoracic endoprosthesis instructions for use state ¿complications associated with the use of the gore® tag® thoracic endoprosthesis may include, but are not limited to, aneurysm enlargement and reoperation.¿ w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2011, the patient underwent endovascular treatment of a thoracic aortic aneurysm using a gore® tag® thoracic endoprosthesis.On an unknown date, a follow-up examination reportedly revealed aneurysm enlargement (amount unknown).No endoleak was observed during the examination.It was reported the patient also developed occasional hemoptysis.On (b)(6) 2022, a reintervention was performed.An intra-operative angiograph showed no endoleak.Two gore® tag® conformable thoracic stent grafts with active control system (ctag-ac) were implanted as proximal and distal extension of the existing gore® tag® thoracic endoprosthesis.After touch-up ballooning was performed, an angiograph reportedly revealed the proximal ctag-ac moved distally about 2cm upon deployment.The cause of the alleged device movement is unknown, and no issues were reported with handling and operating the device per the instructions for use.According to the report, the proximal ctag-ac was positioned at nearly same level as the proximal end of the existing gore® tag® thoracic endoprosthesis.No additional treatment was performed, and the physician will continue to monitor the patient.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: code c19 ¿ a review of the manufacturing records for the device verified that the lot met all pre-release specifications.H6: conclusion coding updated.It should be noted the gore® tag® thoracic endoprosthesis instructions for use state ¿complications associated with the use of the gore® tag® thoracic endoprosthesis may include, but are not limited to, aneurysm enlargement and reoperation.¿.
|
| |
|