| Model Number |
H7493941848250 |
| Medical Device Problem Codes |
Break (1069); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the not very tortuous left anterior descending artery.A 2.50 x 48mm synergy xd drug-eluting stent (des) was advanced but failed to cross the lesion.The device was removed intact, and upon second attempt to cross the device to the lesion, the device fractured.The device was removed and the procedure was completed a 2.25 x 28 synergy des.There were no patient complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: synergy xd mr us 2.50 x 48mm stent delivery system (sds) was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The stent outer diameter was measured using snap gauge and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks and a break 21 cm distal to the strain relief.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 80% stenosed target lesion was located in the not very tortuous left anterior descending artery.A 2.50 x 48mm synergy xd drug-eluting stent (des) was advanced but failed to cross the lesion.The device was removed intact, and upon second attempt to cross the device to the lesion, the device fractured.The device was removed and the procedure was completed a 2.25 x 28 synergy des.There were no patient complications reported.
|
| |
|
Search Alerts/Recalls
|