Model Number DSF1433
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 07/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device was discarded at the facility and, therefore, was not available to gore for direct analysis.It should be noted the gore® dryseal flex introducer sheath instructions for use (ifu) state ¿adequate vessel access is required to introduce the sheath into the vasculature.Careful evaluation of vessel size, anatomy, tortuosity, and disease state (including calcification, plaque, and thrombus) is required to ensure successful sheath introduction and subsequent withdrawal.If vessel is not adequate for access, major bleeding, vessel damage, or serious injury to the patient, including death, may result.¿ w.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2022, a patient underwent endovascular treatment of a left common iliac artery aneurysm using gore® excluder® aaa endoprostheses and a gore® dryseal flex introducer sheath (dsf).During the procedure, an attempt was made to embolize the left internal iliac artery from left-sided access.However, cannulation into the left internal iliac artery was reportedly unsuccessful.The patient¿s right side was accessed, and the left internal iliac artery was successfully embolized.A 14 fr dsf was then inserted from the left side.According to the report, the patient¿s vasculature was extremely tortuous.While advancing the dsf through left external iliac artery, a marked decrease in blood pressure was noted, and the dsf was reportedly outside the vessel.The physician advanced and inflated a molding and occlusion balloon (mob) from the left side to block blood flow at the common iliac artery.Another mob was advanced from the right side and inflated to block blood flow at the distal aorta.The first mob was then deflated, and three stent grafts were implanted as treatment from the left external iliac artery to the middle of the left common iliac artery aneurysm.The procedure went forward to completion, and the patient tolerated the procedure.The physician stated that the left external iliac artery was highly tortuous, and strong resistance was felt when advancing a hard wire.He indicated the access vessel may have been damaged by the wire, and the sheath may have been advanced though a damaged section of the vessel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
damon jackson
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15266345
Report Number3007284313-2022-02083
Device Sequence Number15097475
Product Code DYB
UDI-Device Identifier00733132630004
UDI-Public00733132630004
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/08/2024
Device Model NumberDSF1433
Device Catalogue NumberDSF1433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Initial Report FDA Received Date08/22/2022
Date Device Manufactured05/09/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age84 YR
Patient SexMale
-
-