• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIO-MED DEVICES BIO-MED DEVICES; AIR / OXYGEN BLENDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIO-MED DEVICES BIO-MED DEVICES; AIR / OXYGEN BLENDER Back to Search Results
Model Number 2003FFE5
Medical Device Problem Code Medical Gas Supply Problem (2985)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
The user complained of supply cross-contamination with this blender leaking back to the supply (i.E., air feeding back through the blender and into the oxygen line).No patient injury.
 
Additional Manufacturer Narrative
This blender was built in 2015; it has never been returned to the manufacturer for any factory maintenance, in the 7 years since.We (the manufacturer) have examined the blender sn (b)(4), and root cause has been determined: both input fittings have no rubber duckbill check valve (nor conical filter) inside them.Whoever performed the last service on this unit should not have removed these essential original components, without replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIO-MED DEVICES
Common Device Name
AIR / OXYGEN BLENDER
Manufacturer (Section D)
BIO-MED DEVICES
61 soundview rd
guilford CT 06437
Manufacturer (Section G)
BIO-MED DEVICES
61 soundview rd
guilford CT 06437
Manufacturer Contact
ken close
61 soundview rd
guilford, CT 06437
2034580202
MDR Report Key15266676
Report Number1218704-2022-00002
Device Sequence Number17791992
Product Code BZR
UDI-Device Identifier00813841020256
UDI-Public00813841020256
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K925982
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2003FFE5
Device Catalogue Number2003FFE5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Initial Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/29/2015
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-