| Model Number |
X SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device displayed an "ecg monitoring failure" error message.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The reported problem was observed in the device logs; however the device was put through extensive testing without duplicating the report.The defib receptacle was replaced as a precaution.The device was recertified and returned to the customer.The multifunction cable used was not returned for evaluation at this time.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|