MAUDE Adverse Event Report: ABBOTT VASCULAR HI-TORQUE COMMAND 18 GUIDE WIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR HI-TORQUE COMMAND 18 GUIDE WIRE Back to Search Results
Model Number 1013753
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bradycardia (1751)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that it was a cardiomems procedure.When the cardiomems was being advanced over the hi-torque command 18 guidewire thru the right ventricle, the patient's heart rate dropped to ~25 beats per minute (bpm).It was suggested that the guidewire and/or delivery catheter perturbed the patient's right bundle branch and heart block occurred.Isuprel was given to increase heart rate but did not work.Cardiomems was advanced into the pulmonary artery and deployed successfully.An electrophysiologist came into the room and a temporary pacing wire was placed into the right ventricle, patient was then being paced at 70 bpm.Rather than re-floating the swan into the pulmonary artery past the pacing wire, cardiomems was successfully calibrated to the initial right heart cardiac (rhc) pressures from the pulmonary artery.Patient is stable and was sent to the intensive care unit over the weekend for monitoring due to the heart block instead of being outpatient and ended up receiving a permanent pacemaker.The patient was discharged from the hospital on (b)(6) 2022.There was no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effect of bradycardia is listed in the hi-torque command 18 guide wire for pta instructions for use as a potential adverse event associated with use of this device.A conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.However, the treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE COMMAND 18 GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15267397
Report Number2024168-2022-08984
Device Sequence Number6576803
Product Code DQX
UDI-Device Identifier08717648211980
UDI-Public08717648211980
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K152404
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model Number1013753
Device Catalogue Number1013753
Device Lot Number1120861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2022
Initial Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARDIOMEMS SYSTEM
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age65 YR
Patient SexFemale
Patient Weight106 KG
-
-