| Model Number |
0998-UC-3023-83 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) the cs300 intra-aortic balloon pump (iabp) unit failed battery run time.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the cs300 intra-aortic balloon pump (iabp) failed battery run time during pm.There was no patient involved or adverse event reported.No more details provided.The a getinge field service engineer (fse) replaced lead acid batteries.Completed preventive maintenance.The unit returned fro clinical use.
|
| |
|