• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-330-10-K
Medical Device Problem Codes Signal Artifact/Noise (1036); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The explanted product was not returned to neuropace for analysis.
 
Event or Problem Description
On (b)(6) 2022, during an ecog review, the patient's lead showed evidence of a lead fracture.On (b)(6) 2022 the lead was replaced.There is no information suggesting the cause of the break.The lead was secured with dog bone with lead protection and had been implanted for 1.4 years prior to explant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Common Device Name
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key15271658
Report Number3004426659-2022-00026
Device Sequence Number12385755
Product Code PFN
UDI-Device Identifier00855547005014
UDI-Public010085554700501417220823
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDL-330-10-K
Device Catalogue Number1007607
Device Lot Number28803-1-1-1
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Initial Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age39 YR
Patient SexMale
-
-