| Model Number |
SXPP1A402 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained via: was there any adverse consequence associated with the patient? we confirmed there was no patient consequence in this case.Please provide the lot number: unkonwn.Device return status: we regularly contact with sales rep about the device returning.No further information will be provided.Trade name - irgacare® active ingredient(s) ¿ triclosan dosage form ¿ suture/solid/parenteral strength ¿ = 2360 g/m.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a laparoscopic colorectal cancer surgery procedure on (b)(6) 2022 and barbed suture was used.During the procedure, the fixation tab was broken during continuous suturing.After the needle was passed through the fascia and hanging the fixation tab on the wound, when a surgeon tried to put the 2nd stitch, a joint part between the fixation tab and the suture detached.There was no tension that should normally occur when the fixation tab was applied to the wound, and it detached easily.A demonstration machine was used to demonstrate the tensile strength during suture pulling.The surgeon demonstrated with a needle holder.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 9/28/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h4, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that it was received one open sample pertain to the product code sxpp1a402.During the visual assessment of the needle-suture, it was noted that the swage and attachment area were as expected.Marks on the body needle that appears to be by a surgical instrument could be observed.The suture was examined, body fluids and damages at some barbs were noted.The sides of fixation tab were observed to be broken.It should be noted that a 100% inspection is perform via automotive vision system to ensure the tab is present and complete during manufacturing.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name - irgacare®.Active ingredient(s) ¿ triclosan.Dosage form ¿ suture/solid/parenteral.Strength ¿ = 2360 g/m.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc-(b)(4).Date sent to the fda: 9/8/2022 this is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The following additional information was received: lot number : unk => rlmjxt trade name - irgacare® active ingredient(s) ¿ triclosan dosage form ¿ suture/solid/parenteral strength ¿ = 2360 ¿g/m.
|
| |
|