MAUDE Adverse Event Report: AROA BIOSURGERY LTD OVITEX; OVITEX ABSORBABLE MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AROA BIOSURGERY LTD OVITEX; OVITEX ABSORBABLE MESH Back to Search Results
Model Number F10246-1012G
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Dysphagia/ Odynophagia (1815); Scar Tissue (2060); Obstruction/Occlusion (2422); Device Embedded In Tissue or Plaque (3165)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Insufficient device information received to conduct investigation of manufacturing records.A supplemental report will be filed should additional information become available.
 
Event or Problem Description
A patient self-reported complications after hiatal hernia repair with a tif procedure in february 2020 with ovitex 1sr.The patient reported experiencing dysphagia for approximately 10 months and underwent additional surgery (time unknown).The patient reported potential device embedment and scar tissue at the esophagus.
 
Additional Manufacturer Narrative
An investigation update is provided below: upon review of medical records provided by the patient, it is unlikely that the ovitex device caused or contributed to the issues noted in the original report.The patient reported a complex medical history prior to hernia repair with ovitex.The initial procedure was also conducted together with a tif procedure that implanted multiple permanent fasteners.Multiple endoscopies showed no injuries to the esophagus and that the tif valve was intact (4 and 6 months after initial surgery, respectively).During a (b)(6) 2020 (10 months post-op) endoscopy, an extruded suture was removed but no extruded mesh was noted.In (b)(6) 2021, (17 months after initial surgery) the tif was attempted to be reversed.No records are available after this time.A batch record review of the lot utilized in the original procedure showed no non-conformances or anomalies that may have caused or contributed to the issue noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OVITEX
Common Device Name
OVITEX ABSORBABLE MESH
Manufacturer (Section D)
AROA BIOSURGERY LTD
2 kingsford smith place,
airport oaks
auckland, 2022
NZ  2022
Manufacturer (Section G)
AROA BIOSURGERY LTD
2 kingsford smith place,
airport oaks
auckland, 2022
NZ   2022
Manufacturer Contact
tina o'brien
2 kingsford smith place,
airport oaks
auckland, 2022
NZ   2022
MDR Report Key15271915
Report Number3007321028-2022-00028
Device Sequence Number6475727
Product Code FTM
UDI-Device Identifier09421904065307
UDI-Public09421904065307
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date07/31/2020
Device Model NumberF10246-1012G
Device Catalogue NumberF10246-1012G
Device Lot NumberERT-9B10
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer07/28/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date03/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/18/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TIF PROCEDURE WITH SUTURE FASTENERS
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-