Model Number 1012535-40
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the absolute pro vascular stent delivery system was removed from the hoop without difficulty.The stylet was removed without difficulty.It was noted that the stent was exposed.The device was not used and there was no patient involvement.There was no patient involvement and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported premature activation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that during unpackaging and/or during preparation for use inadvertent mishandling resulted in the reported premature activation; however this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABSOLUTE PRO VASCULAR SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15272172
Report Number2024168-2022-09034
Device Sequence Number6476255
Product Code NIP
UDI-Device Identifier08717648176067
UDI-Public08717648176067
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P110028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number1012535-40
Device Catalogue Number1012535-40
Device Lot Number9112061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/01/2022
Initial Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-