• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. TRULIANT KNEE COMPONENTS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. TRULIANT KNEE COMPONENTS; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number TRULIANT TIB IMP PSC INSERT SZ 4, 17MM
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Swelling/ Edema (4577)
Date of Event 11/08/2019
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, this patient had a revision due to the capsule releasing, causing pain and swelling in the right tka site.This event report was received through clinical data collection activities.
 
Additional Manufacturer Narrative
Pending evaluation.Concomitant medical products: 02-020-11-0340 - truliant ps cem fem ps cem right sz 4 sn: (b)(4), 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t sn: (b)(4), 200-07-35 - advanced patella 35mm 3 peg implant sn: (b)(4).
 
Additional Manufacturer Narrative
Section h10: (h3) based on review of all available information, there is no evidence to suggest that the reported event is related to any design issues.The cause of the pain and subsequent revision cannot be conclusively determined; however, it is most likely related to the patient¿s underlying condition.H7: z-0023-2022.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b3, d1, d4 (remove all from initial report-suspect device unknown), d8, g4 (remove 501k from initial report-unknown) h4 (remove manufacture date from initial report-unknown), h6 clinical code, component code, investigation findings, investigation conclusion codes.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT KNEE COMPONENTS
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key15272224
Report Number1038671-2022-00949
Device Sequence Number17589101
Product Code JWH
UDI-Device Identifier10885862305336
UDI-Public10885862305336
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2026
Device Model NumberTRULIANT TIB IMP PSC INSERT SZ 4, 17MM
Device Catalogue Number02-022-44-4017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2022
Supplement Date Received by Manufacturer08/02/2022
01/14/2026
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date09/23/2022
01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age62 YR
Patient SexMale
Patient Weight125 KG
-
-