MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION COYOTE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24706
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left anterior tibial artery.A 3.5mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 7 atmospheres, the balloon ruptured.After a 1.75mm rotablation was performed and the procedure was completed.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 39mm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed there is a pinhole in the balloon.
 
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in the left anterior tibial artery.A 3.5mm x 220mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 7 atmospheres, the balloon ruptured.After a 1.75mm rotablation was performed and the procedure was completed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COYOTE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15272454
Report Number2124215-2022-31886
Device Sequence Number14922734
Product Code DQY
UDI-Device Identifier08714729797326
UDI-Public08714729797326
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K111295
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2023
Device Model Number24706
Device Catalogue Number24706
Device Lot Number0026520045
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date08/22/2022
Supplement Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age76 YR
Patient SexFemale
Patient Weight75 KG
-
-