• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 502-11-50D
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the patient's left hip was revised due to pain.A shell, 2 screws, liner and head were revised.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.The following devices were also listed in this report: device name# trident 0 deg x3 insert 32mm head; cat#: 623-00-32d; lot#: mmd9yp; device name# 32mm +4 v40 taper vit head; cat#: 6260-5-232; lot#: 42583403; device name# 6.5 cancellous bone screw 25mm; cat#: 2030-6525-1; lot#: mma924; device name# 6.5 cancellous bone screw 35mm; cat#: 2030-6535-1; lot#: mmanh7.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15273482
Report Number0002249697-2022-01215
Device Sequence Number16808908
Product Code LPH
UDI-Device Identifier04546540432025
UDI-Public04546540432025
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K013676
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2017
Device Model Number502-11-50D
Device Catalogue Number502-11-50D
Device Lot Number41192601
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/27/2022
Initial Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age89 YR
Patient SexFemale
Patient Weight65 KG
-
-