• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0; TUBE, TRACHEAL (W/WO CONNECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0; TUBE, TRACHEAL (W/WO CONNECTOR Back to Search Results
Model Number IPN044799
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A device history record review was performed, and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event or Problem Description
It was reported that: the doctor found cuff leakage when doing clinical setting before using on patient.No impact on patient.
 
Additional Manufacturer Narrative
Qn#(b)(4).The customer returned one unit 5-12614 et tube, cuffed, spiral-flex 7.0 for investigation.The et tube was visually examined with and without magnification.Visual examination of the returned device revealed that there were multiple holes in the balloon cuff which would prevent the cuff from inflating.The ifu for this product states, "the tracheal tube cuff inflation system, including the valve, should be checked prior to intubation.If a malfunction is a detected in any part of the system, the tube should not be used." "care should be taken to avoid damaging the thin-walled cuffs during intubation.If cuff is damaged, the tube should not be used." the ifu also states, "cuff pressure should be monitored routinely to assure that an adequate tracheal seal is maintained and that the cuff has not become over-inflated.The tracheal tube cuff should be slowly inflated with the minimum amount of air required to provide an effective tracheal seal.Do not inflate with a measured volume of air or by "feel" of pressure in the syringe since little resistance should be felt during inflation." the reported complaint was confirmed based upon the sample received.The returned et tube was found to have multiple holes in the balloon cuff which would prevent it from being able to stay inflated.A device history record review was performed with no evidence to suggest a manufacturing related cause.All et tubes are 100% inspected for inflation and deflation at the time of manufacturing.It is unlikely that this type of damage was present at the time of manufacturing.Therefore , based upon the damage observed, unintentional user error caused or contributed to this event.
 
Event or Problem Description
It was reported that: "the doctor found cuff leakage when doing clinical setting before using on patient.No impact on patient.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON ET TUBE, CUFFED, SPIRAL-FLEX 7.0
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15274530
Report Number3003898360-2022-00398
Device Sequence Number2595571
Product Code BTR
UDI-Device Identifier04026704397402
UDI-Public04026704397402
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K844296
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date08/24/2025
Device Model NumberIPN044799
Device Catalogue Number5-12614
Device Lot Number73H2000543
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer09/14/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-