• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH DRILL BIT AXSOS 3 LOCKING, SHORT, Ø3.1 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH DRILL BIT AXSOS 3 LOCKING, SHORT, Ø3.1 X 216MM; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 705031S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 07/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a drill bit broke and the drill tip remained in the patient.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event or Problem Description
It was reported that a drill bit broke and the drill tip remained in the patient.
 
Additional Manufacturer Narrative
The reported event could be confirmed, since the drill bit is broken as complained.The device inspection revealed the following: the identification of the device drill bit axsos 3 ti locking, short, ø3.1 x 216mm is confirmed based on the lot number and the catalog number marked.The visual inspection has shown that in the front part of the drill bit, there are strong stress marks above the fracture visible.Microscope pictures shows lateral view of the fracture zone, the edge is at the fracture face deformed, which indicates a metallic contact in this area.The breakage surface reveals typical characteristics of a brittle fracture i.E.Relatively smoother surface at the center than at the edges.Since the breakage was brittle (sudden) in nature, there is evidence of abrupt features especially by plateau on both sides of the fracture face.Furthermore, a hardness test according to rockwell on the returned item indicates the material being in accordance to the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to a user related issue.The failure was caused by too much force had been applied.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT AXSOS 3 LOCKING, SHORT, Ø3.1 X 216MM
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ   2545
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15275911
Report Number0008031020-2022-00410
Device Sequence Number12383867
Product Code HWE
UDI-Device Identifier07613327083330
UDI-Public07613327083330
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number705031S
Device Catalogue Number705031S
Device Lot Number25129H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer09/19/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date10/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-