| Model Number |
1012451-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat an unspecified lesion with heavy calcification and mild tortuosity.The 3.5x12 mm nc trek balloon dilatation catheter was advanced with resistance noted with the anatomy.The balloon was inflated twice and ruptured at 15 atmospheres (atm).There were no adverse patient effects and there was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|