• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL X1 STRAIGHT 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL X1 STRAIGHT 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLX137S
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 07/19/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Date sent: 8/23/2022.Batch #: x94z8k.Additional information was requested and the following was obtained: what was the indication for surgery? low anterior resection for cancerous tumor.What was the patient's pre-op anti-coagulation therapy? normal.Were any anti-coagulation drugs given prior to the procedure? no.Did the patient have any anti-coagulopathy disorders? no, it was a vessel sealing isssue.The device did not seal/transect the vessel completely.Had the patient received any radiation or chemo therapy prior to the operation? no.What was the approximate size of the ima? 5mm.Was the ima completely skeletonized when taken? yes, per the surgeon and fellow.Where within the triple burn was the vessel transected? in the middle burn.Was the completion tone heard on all activations? yes.Was the device fully latched for each of the triple burns? yes.Did the patient require any blood transfusions? no.What is the patient's current status? recovering.Is the surgeon interested in speaking with ethicon medical and engineering staff? potentially.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during an lar the surgeon mobilized around the ima.He then triple burned on the ima then transected across to seal off the vessel.It did not seal.There was significant bleeding.He then had to open the patient and complete the case in open status.There was a delay of 2 hours.The patient lost 600 cc's of blood.The patient is currently recovering.
 
Additional Manufacturer Narrative
(b)(4).Date sent: 9/16/2022.D4 batch #: x94z8k.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with no apparent damage.The device was tested on the generator and passed all functional testing.The energy output delivered from the device was verified.All tones were heard during functional testing.The device was tested with the test media and no anomalies were found.There were no anomalies noted with the functionality of the device.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reach on the cause of the reported event.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 STRAIGHT 37CM SHAFT
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
JABIL (TAJUANA)
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
329348013
MDR Report Key15278024
Report Number3005075853-2022-05509
Device Sequence Number6485597
Product Code GEI
UDI-Device Identifier10705036015482
UDI-Public10705036015482
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNSLX137S
Device Catalogue NumberNSLX137S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2022
Supplement Date Received by Manufacturer08/23/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-