• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. JELCO PROTECTIV SAFETY IV CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. JELCO PROTECTIV SAFETY IV CATHETERS; CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM Back to Search Results
Model Number 306301
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.Two (2) product sample photos were received for evaluation.Visual inspection found the photos taken of the complaint sample did not include magnified closed up views of the catheter bevel tip for evidence of mechanical witness marks or manufacturing damage or of the needle guard assembly for evidence of cannula bevel tip irregularities, excess adhesive, mechanical witness marks or manufacturing damage.The second photo provided the affected lot number.No functional testing could be performed as no samples were received.The reported issue could not be verified or confirmed per pictures provided for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Event or Problem Description
It was reported that there were multiple reports of catheter sheering upon insertion over the last few months.No patient injury and no return product was additionally reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JELCO PROTECTIV SAFETY IV CATHETERS
Common Device Name
CATHETER, INTRAVASCULAR, THERAPEUTIC, SHORT-TERM
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
201 west queen st.
southington CT 06489
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15278286
Report Number3012307300-2022-15923
Device Sequence Number6439119
Product Code FOZ
UDI-Device Identifier10351688071224
UDI-Public10351688071224
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K990236
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/18/2024
Device Model Number306301
Device Catalogue Number306301
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2022
Initial Report FDA Received Date08/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-