MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CFS22, 12X100 KII SLEEVE ADVFIX 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CFS22, 12X100 KII SLEEVE ADVFIX 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number CFS22
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
Procedure performed: unknown.Event description: while inserting the trocar (without obturator) through minilap incision, pieces of plastic fell into the abdomen.The staff is unsure if all the fragments have been secured.Patient status: no patient injury or illness occur associated with the complaint event.
 
Additional Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the complainant¿s experience of a damaged cannula tip, the cannula tip contained a crack and the glue line was fragmented.Based on the condition of the returned unit, it is possible that the cracked cannula tip and fragmented glue line were caused by an object or instrument coming in contact with the cannula tip and glue line during the procedure.It is also possible that the cannula and glue line were more susceptible to fragmentation.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event or Problem Description
Procedure performed: unknown.Event description: while inserting the trocar (without obturator) through minilap incision, pieces of plastic fell into the abdomen.The staff is unsure if all the fragments have been secured.Patient status: no patient injury or illness occur associated with the complaint event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CFS22, 12X100 KII SLEEVE ADVFIX 12/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
aaron fulcher
22872 avenida empresa
rancho santa margarita, CA 92688
9497135765
MDR Report Key15279965
Report Number2027111-2022-00712
Device Sequence Number2292285
Product Code GCJ
UDI-Device Identifier00607915117740
UDI-Public(01)00607915117740(17)250523(30)01(10)1449925
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K083638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/23/2025
Device Model NumberCFS22
Device Catalogue Number101472330
Device Lot Number1449925
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2022
Supplement Date Received by Manufacturer07/27/2022
Initial Report FDA Received Date08/23/2022
Supplement Report FDA Received Date10/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-